This section
Dichroic mirrorBased on the development of fluorescent light cluster imaging, its main optical parameters achieve high reflection in the 350-510nm band and high transmission in the 530nm-1000nm band. This dichroic mirror is designed for 45 degree incidence, with a small transition band between high reflection and high transmission bands, making it suitable for most applications that require fluorescence detection.
Technical Parameters:
Process: Ion assisted dura mater;
Material: k9/b270 supports customization of any material;
Wavelength range: 200nm~1100nm;
Incident angle: 45 degrees;
Peak transmittance (T peak):>92%;
Blocking depth: OD4;
Starting wavelength (lambda; cut on): 520nm;
Dimension: 50 * 50 * 1mm can be customized according to customer requirements, with fast delivery time;
Surface quality: 60/40;
Surface parallelism:λ /4;
Application: Machine vision imaging, fluorescence microscopy, PCR fluorescence quantitative analyzer, fluorescein, fluorescent dyes, fluorescent chromophore detection.
Actual measured spectral curve: